GPR-2800 IS GPR-2800 IS-LD % OXYGEN TRANSMITTE氧...
- 型号:GPR-2800 IS
- 供应商:托驰(上海)工业传感器有限公司
- 供应商报价:面议
- 标签:GPR-2800 IS-LD % OXYGEN TRANSMITTE氧气分析仪,-1,托驰(上海)工业传感器有限公司
ATEX and cUL certified % oxygen transmitter measures O2. Features certification to ATEX Directive 94/9/EC and cUL designed exclusively for NATURAL GAS APPLICATIONS. This analyzer features an advanced galvanic trace PPM oxygen sensor, a 18 month operating sensor life, menu driven controls, and two (2) field selectable alarm set points. Pictured with optional A-3393 modular sample system designed to remove moderate amounts of free liquids and H2S from natural gas streams.
Certifications:
cUL Class I Div I Group C, D
ATEX Directive 94/9/ II 2G
Ex d [ib] ib IIB T4
Tamb -10oC to 50oC
| *Accuracy: | < 2% of FS range under constant conditions | Analysis: | 0-1%, 0-5%, 0-10%, 0-25% FS ranges; auto-ranging or fixed single range |
| Application: | Oxygen analysis in inert, hydrocarbon, hydrogen, mixed and acid (CO2) gas streams | Approvals: | ATEX Directive 94/9/EC II 2 G Ex d [ib] ib IIB T4 Tamb -20?C to +50?C |
| Area Classification: | cUL for Class 1, Division 1, Group C, D hazardous | Alarms: | None |
| Calibration: | Air or certified span gas of O2 balance N2 | Compensation: | Barometric pressure and temperature |
| Connections: | 1/4″ compression tube fittings | Controls: | Water resistant keypad; menu driven range selection, calibration and system functions |
| Display: | Graphical LCD 2.75” x 1.375”; resolution 0.001%; displays real time ambient temperature and pressure | Enclosure: | NEMA Type 3R suitable for rain in outdoor applications |
| Flow: | Not flow sensitive; 1-2 SCFH recommended | Linearity: | + 1% of FS |
| Pressure: | Inlet – regulate to 5-30 psig to deliver 1-2 SCFH flow to transmitter; vent – atmospheric | Power: | 12-24 VDC 2 wire loop power |
| Response Time: | 90% of final reading in < 2 minutes | Sample System: | Unique liquid drain sensor manifold, flow indicator |
| Sensitivity: | < 0.5% of FS range | Sensor Model: | GPR-12-333-LD for inert gases |
| Sensor Life: | 32 months in air at 25ºC and 1 atm | Signal Output: | 4-20mA |
| Operating Range: | Recommended -10ºto 45ºC | Warranty: | 12 months analyzer; 12 months sensor |
| Wetted Parts: | Stainless steel |